In what order should I troubleshoot unstable volume?

2026.08.08

Work from the outside in, simplest first:

1) Supply pressure: fluctuation shows up directly as volume variation; check the regulator and reservoir level first.

2) Needle: wear, deformation or partial clogging drifts the flow; swap in an identical needle to compare.

3) Bubbles and residue: incomplete de-gassing or cured residue in the flow path causes intermittent dispensing.

4) Stator and seals: wear after long service causes metering deviation - a normal consumable replacement.

If it is still unstable, send us your volume test records and we will help analyse.